- Overview
- Recommended Products
Perfect size of 0.6mm slice thickness detector
Advance iterative reconstruction based on the noise model.
3.5M metal high capacity X-ray tube
350mA tube current selection
32 slice fast and precision scan
70kV advanced low dose scan
Efficiency from positioning to post-processing.
High definition imaging(20 Lp/cm)with minimum dose.
Easy to use with large 76 cm scan bore and 250kg patient couch.
32-row submillimeter detector with 20mm coverage realize fast scan and high quality images
5.3MHU/8.0MHU tube with 50kW/80kW generator provide long life and high profit for hospitals.
More Clarity with Less Dose
76cm large scan space provides greater openness for easier patient position.
64 Rows/128 slice imaging platform generates full applications
Fast workflow and excellent visualization solution
Significant dose reduction and excellent clinical performance
Cardiac application package
System |
Width |
Length |
Height |
Weight |
Mm |
Mm |
Mm |
Kg |
|
Gantry |
1970 |
940 |
1850 |
1200 |
Table |
580 |
2400 |
950 |
350 |
Console |
400 |
630 |
630 |
50 |
Distributor |
810 |
435 |
665 |
200 |
Working Environment & Siting Requirement
Temperature
Scan room:20 ~ 26 ℃ Operation room:18 ~ 28 ℃
Humidity of Scanning room :
Scan room 30% -70%,Control room 20%- 80%
Atmospheric Pressure : 70 kPa ~ 106 kPa
Power Supply Requirements:
Power Capacity :50 kVA
Input Voltage :Equipped with transformer,
Power supply option: 380-480VAC
Voltage Variation : Tolerance < ±5%
Frequency :50Hz/60Hz±1Hz