- Conspectus
- Producta Commendata
Model: HDR50A

Compositio Systematis
| Systema Digitalis Radiographiae Altae Frequentiae HDR50A | |
| Altus frequentia altus voltus radiogenitor x | V |
| Constitutio Tubuli X-ray | Made-in-China |
| Collimator | A-VM |
| Detectōr digitālis | Branda Iray Sinica |
| DR Machinae Framen | S-V |
| Tabula Radiographica | S-V |
| Computātrum et LCD | DCSM1F/ML2002 |
| Cavum Altius Voltaticum | 75kV (longitudo: 8m) |
| Reticulum (Genus:JPI) | 100cm et 180cm Made-in-Korea |
| O-IV cum consolio operationis | |
| Potentia | AC380V ± 38V |
| Tensio Tubi Radiographici | 40kV ~ 150kV |
| Currentia Tubi Radiographici | 10mA ~ 630mA |
| Tempus | 2ms ~ 6300ms |
| Resistentia Interna Permissa | 380V: 0,5Ω |
| Variatio Frequentiae Permissa | 50Hz/60Hz ± 1Hz |
| Opus Frequency | ≥300kHz |

Constitutio Tubuli X-ray
| Focus | 0.6mm/1.2mm |
| Mobilitate rotating | 2800rpm |
| Tensio Tubi Peek | 150kV |
| Parvus focus | 20kw |
| Magnus focus | 50kW |
Collimator M-38
| Modus Regulandi | Manualis |
| Potentia | 24VAC |
| Filtratio Innata | 1.2mmAL |
Detector Digitalis Rayence
| Area Scandendi | 17 × 17 pollices |
| Loci resolutio | non minus quam 3,6lp/mm |
| Matrix Pixel Totus | 3,072(h) × 3,072(v) |
| Distantia Pexel | 139 |
| Tempus Praevisionis Imaginis | 3s |

DR Machinae Structura YS-V
| Ampitudo motus horizontalis lectuli | ≥250mm |
| Amplitudo motus verticalis lectuli | ≥825mm |
| Magnitudo tabulae (Longitudo × Latitudo × Altitudo) | 2100 × 825× 655mm |
| Modus Operationis | Manuale vel electronicum |
| Onus-afferentem | Minus quam 250kg |
| CPU | ≥2.8GHz |
| Memoria | ≥4GB |
| Capacitas Disci | ≥500GB |
| LCD Medicale | 23 pollices |
Praecipua lineamenta:
1. Generator Altae Frequentiae
1.1 Functiones: Radiographia digitalis X-radiationis, radiographia cum filtro X-radiationis, radiographia cum stative Bucky, ARP et AEC.
1.2 Vis electrica nominalis: 50 kW
1.3 Tensio tubi: 40 kV ~ 150 kV
1.4 Currentia tubi: 10 mA ~ 630 mA
1.5 Tempus: 2 ms ~ 10 s
1.6 mAs: 0.1mA·s~630mA·s
1.7 Frequencia operans: ≥400 kHz
1.8 Tabula directionis: LCD cum tactili scuto
2. Tubus X-radiationis
2.1 Tensio maxima tubi: 150 kV
2.2 Focus: Focus parvus 0.6mm; Focus magnus: 1.2mm
2.3 Potentia inducta in discum: focus parvus 19kW, focus magnus 50kW
2.4 Angulus Anodi: 12°
2.5 RPM Anodi: 2800rpm
2.6 Capacitas caloris: 150 kHU
3. Detector Planus
3.1 CsI
3.2 Magnitudo: 17×17pollicum
3.3 Pixel: 139μm
3.4 Matrices pixelarum: 3072×3072
3.5 conversio A/D: 16Bit
4. Mensa Radiographica
4.1 Magnitudo: 2100mm×825mm×655mm (L×L×A)
4.2 Campum motus horizontalis: ≥250 mm;
4.3 Campi motus tabulae: ≥250mm;
4.4 Clavis longitudinalis: Clavis electromagnetica, vis frenandi ≥100N, cum commutatore pedali, flexibilis et commodus.
4.5 Cum motu mechanico electrico vel manu regendo, et functione automatica sequendi et dirigendi coniuncta cum tubo radiographico et detectore.
4.6 Magnitudo cancelli: ≥15"×18"; Distantia focalis cancelli: f0=100cm; Rationem cancelli: r=10:1; Densitas cancelli: N=103Lineae/Inch4.7 Columna Tubi
4.7.1 Structura firma et solida
4.7.2 Rotatio columnae: ±180°
4.7.3 Campi motus secundum tabulam: ≥1575 mm, potest in quovis puncto concludi
4.7.4 Tubus sursum et deorsum movendus: 930 mm ± 10 mm, potest in quovis puncto concludi
4.7.5 Rotatio tubi: ±180°
4.7.6 Distantia maxima a foco tubi ad pelliculam: ≥1200 mm.
5. Bucky Stans
5.1 Campi motus sursum et deorsum: ≥1200 mm;
5.2 Ad altitudinem maximam, distantia inter solum et centrum capsulae est ≥1500mm;
5.3 Magnitudo cancelli: 18×18", focus cancelli: f0=180cm; ratio cancelli: r=10:1; densitas cancelli: N=103Lineae/Poll.
6. Colimator: Flexibilis et commodus ad operandum, et omnis area radiographica rectangularis disponibilis est.
| Non. | Nomina Partium | QTY | Notae | |
|---|---|---|---|---|
| 1 | Generator Altus Frequentia | 1 PC | ||
| 2 | Tabula Imperii | 1 PC | ||
| 3 | Tubus Radiologicus | 1 Set | ||
| 4 | Collimator | 1 PC | ||
| 5 | Detector Planus cum Programmate | 1 Set | 17"*17 | |
| 6 | Bucky Stand | 1 PC | ||
| 7 | Tabula Radiographica | 1 PC | ||
| 8 | Statio Laboris (Computer) | 1 Set | -- | |
| 9 | Cancellus (optionale) | 2 pc | 18"×18" | JPI |
| 10 | Cavum Altius Voltaticum | 1 par | 75kV |