- Ko'rib chiqish
- Tavsiya etilgan mahsulotlar
Model: HDR50A

Tizim konfiguratsiyasi
| HDR50A Raqamli Yuqori Chastotali Rentgenologik Tizim | |
| Yuqori chastota yuqori voltaj x-ray generatori | V |
| X-ray Tube Assembly | Xitoyda ishlab chiqarilgan |
| Kollimator | A-VM |
| Raqamli detektor | Xitoy Iray brendi |
| DR Mashinasi Asosiy Qisimi | S-V |
| Rentgen stoli | S-V |
| Kompyuter va LCD | DCSM1F/ML2002 |
| Yuqori kuchlanish kabeli | 75kV (uzunligi: 8m) |
| Tarmoq (Brend: JPI) | 100 sm va 180 sm O'rkashapeydachi |
| Boshqaruv pulti bilan O-IV | |
| Quvvat | AC380V ± 38V |
| Rentgen nuri trubkasi kuchlanishi | 40kV ~ 150kV |
| Rentgen nuri trubkasi toki | 10mA ~ 630mA |
| Vaqt | 2ms ~ 6300ms |
| Ruxsat etilgan ichki qarshilik | 380V: 0,5Ω |
| Ruxsat etilgan chastota o'zgarishi | 50Hz/60Hz ± 1Hz |
| Ish tezligi | ≥300kHz |

X-ray Tube Assembly
| Fokus | 0.6mm/1.2mm |
| Aylanma tezligi | 2800rpm |
| Peek trubkasi kuchlanishi | 150kV |
| Kichik fokus | 20KW |
| Katta fokus | 50kW |
Kolimator M-38
| Sozlash usuli | Qo'llanma |
| Quvvat | 24VAC |
| Xususiy filtratsiya | 1.2mmAL |
Raents raqamli detektori
| Skanerlash maydoni | 17 × 17 dyuym |
| Orinli yechim | 3,6 lp/mm dan kam bo'lmasa |
| Umumiy piksel matritsasi | 3,072(g) × 3,072(v) |
| Piksel pichi | 139 |
| Rasmni ko'rish vaqti | 3S |

DR Mashina Ramkasi YS-V
| Stolning gorizontal harakat doirasi | ≥250mm |
| Stolning vertikal harakat doirasi | ≥825mm |
| Stol yuzasining o'lchami (uzunlik × kenglik × balandlik) | 2100 × 825× 655mm |
| Ishlash tartibi | Qo'lda yoki elektron |
| YUK-TASHISH | 250 kg dan kam |
| CPU | ≥2.8GHz |
| Xotira | ≥4GB |
| Disk hajmi | ≥500GB |
| Tibbiy LCD | 23 inch |
Asosiy xususiyatlari:
1. Yuqori chastotali generator
1.1 Funktsiyalar: Raqamli rentgenografiya, rentgen filtri orqali rentgenografiya, Buki tayanchi orqali rentgenografiya, ARP va AEC.
1.2 Nominal elektr quvvati: 50 kW
1.3 Nomiyl vol't: 40 kV~150 kV
1.4 Nomiyl tok: 10 mA~630 mA
1.5 Vaqt: 2 ms~10 s
1.6 mAs: 0.1mA·s~630mA·s
1.7 Ish chastotasi: ≥400 kHz
1.8 Boshqaruv paneli: LCD sensorli ekran
2. Rentgen nuri trubkasi
2.1 Maks. trubka kuchlanishi: 150 kV
2.2 Fokus: Kichik fokus 0.6mm; Katta fokus: 1.2mm
2.3 Nishonga kiritiladigan quvvat: kichik fokus uchun 19kW, katta fokus uchun 50kW
2.4 Anod burchagi: 12°
2.5 Anod aylanish tezligi: 2800rpm
2.6 Issiqlik sig'imi: 150kHU
3. Tekis panel detektori
3.1 CsI
3.2 O'lchami: 17×17 dyuym
3.3 Piksel: 139 μm
3.4 Piksel matritsasi: 3072×3072
3.5 A/D aylanishi: 16Bit
4. Rentgen stoli
4.1 O'lchami: 2100 mm × 825 mm × 655 mm (U × B × B)
4.2 Gorizontal harakatlanish doirasi: ≥250mm;
4.3 Stol yuzasining harakatlanish diapazoni: ≥250 mm;
4.4 Uzunlama bloklash: Elektromagnit qulflash, tormoz kuchi ≥100 N, oyoq pedali bilan, moslashtirish uchun qulay va ixcham.
4.5 Mexanik harakatni elektr yoki qo'lda boshqarish imkoniyati mavjud, rentgen nuri trubkasi va detektor bilan avtomatik ravishda tekislash funksiyasiga ega.
4.6 To'rsimon tuzilma hajmi: ≥15"×18"; To'rsimon fokus masofasi: f0=100 sm; To'rsimon nisbati: r=10:1; To'rsimon zichligi: N=103 chiziq/dyuym 4.7 Truba ustuni
4.7.1 Mustahkam va barqaror tuzilish
4.7.2 Ustunning aylanish burchagi: ±180°
4.7.3 Stol bo'ylab harakatlanish doirasi: ≥1575 mm, istalgan nuqtada blokirovka qilish imkoniyati mavjud
4.7.4 Naycha yuqoriga va pastga harakatlanadi: 930 mm ± 10 mm, istalgan nuqtada blokirovka qilish imkoniyati mavjud
4.7.5 Quvur aylanishi: ±180°
4.7.6 Nurlanish manbaidan filmgacha bo'lgan maksimal masofa: ≥1200 mm.
5. Bakkiston tayanchi
5.1 Yuqori-pastga harakatlanish doirasi: ≥1200 mm;
5.2 Eng yuqori balandlikda quti markazi bilan pol o'rtasidagi masofa ≥1500 mm;
5.3 To'rsimon hajm: 18×18", To'rsimon fokus: f0=180 sm; To'rsimon nisbati: r=10:1; To'rsimon zichligi: N=103 chiziq/dyujim
6. Kollimator: Ishlatish uchun mos va qulay, ixtiyoriy to'rtburchakli radiografiya maydoni hajmi mavjud.
| Yo'q. | Qism nomlari | Miktor | Izohlar | |
|---|---|---|---|---|
| 1 | Юксак частоталы генератор | 1 dona | ||
| 2 | Boshqaruv paneli | 1 dona | ||
| 3 | X-ray Tube | 1 to'plam | ||
| 4 | Kollimator | 1 dona | ||
| 5 | Dasturiy ta'minotli tekis panel detektori | 1 to'plam | 17"*17 | |
| 6 | Bucky podstavka | 1 dona | ||
| 7 | Rentgen stoli | 1 dona | ||
| 8 | Ish stansiyasi (Kompyuter) | 1 to'plam | -- | |
| 9 | To'rsimon tuzilma (ixtiyoriy) | 2 dona | 18"×18" | JPI |
| 10 | Yuqori kuchlanish kabeli | 1 ta jufti | 75kV |